Atomic Absorption Spectrophotometer
WFX-210 Atomic Absorption Spectrophotometer
Features for WFX-210 Atomic Absorption Spectrophotometer
Innovated flame analysis technique
- The patented flame analysis technique adopting oxygen-rich air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, Al, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950ºC, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost. Oxygen-rich flame will not pollute the environment and is not harmful to human bodies. It is a break-through in flame AAS analysis.
- Popular Czerny-Turner grating monochromator with 1800l/mm provides higher resolution.
- All plastic pre-mix spray chambers with high anti-corrosion ability.
Accurate and reliable graphite furnace analysis
- Light-controlled graphite furnace power supplier, adopting FUZZY-PID temperature control technique and dual curve working mode, ensures fast heating, accurate and stable temperature control and good temperature reproducibility. Temperature auto-correction function provided.
- Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact.
- High accuracy and reproducibility of the graphite furnace auto-sampler, realizing fully automation of graphite furnace analysis.
Fully automated control system
- Auto-changeover and position optimization of integrated flame/graphite furnace atomizer.
- Automatic 6-lamp turret. High performance HCL can be used directly, which features high intensity spectral line and low background.
- Auto-ignition, automatic spectral bandwidth selection.
- Fully automated wavelength scanning and peaking.
- Adopting mass flowmeter with program control ensures accurate gas intake volume.
- Automatic setting of Gas flow
Perfect safety protection measures
- Alarm and automatic safety protection to fuel gas leakage, air pressure deficiency and abnormal flame extinction
- Spray chamber made of explosion-proof material, all titanium burner and explosion-proof stopper for flame analyses.
- Alarm and protection function to argon pressure deficiency, insufficient cooling water supply, over-heating and over-current ect. In graphite furnace analyses.
Advanced and reliable electronic design
- Adopting large-scale programmable logic array and Inter IC bus technique and highly reliable adapters for electric design.
- European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electric system.
- Standard RS-232 serial interface communication.
Easy and practical analysis software
- Easy-to-use AAS analysis software is made under Windows operating system.
- Automatic curve fitting, automatic sensitivity correction.
- Automatic calculation of sample concentration (content), mean value standard deviation and relative standard deviation calculation.
- Multi-elements determination in sequence to the same sample.
- Measured data and final results can be printed out and edited in Excel format.
Comparison
Characteristic Mass of Some Elements using Oxygen-rich air-C2H2 flame and other flame methods (Unit: ug/ml)
Element |
Wavelength |
Oxygen-rich air-C2H2 flame |
N2O-C2H2 flame |
Air-C2H2 flame |
Ca |
422.7 |
0.009 |
0.05 |
0.07 |
Yb |
378.8 |
0.037 |
0.08 |
7.6 |
Eu |
459.4 |
0.137 |
0.3 |
3.0 |
Al |
309.3 |
0.4 |
0.7 |
|
Sr |
460.7 |
0.016 |
0.1 |
0.15 |
Ba |
553.5 |
0.1 |
0.4 |
10.0 |
Mo |
313.3 |
0.15 |
0.4 |
0.8 |
W |
255.1 |
3.2 |
5.0 |
|
Ga |
287.4 |
0.4 |
1.0 |
1.3 |
Sm |
429.7 |
2.92 |
8.5 |
|
La |
550.1 |
37.2 |
35.0 |
|
Sn |
224.6 |
0.8 |
3.0 |
50 |